Ricoh IM3500 Intermittently Reports SC547 Fault

Visitor 414 0
Symptom: IM3500 intermittently reports SC547 malfunction
Reason: BICU board malfunction

Solution:
IM3500 reports to SC547-01-02-03 and SC552, and can be used normally after fault reset
1. Measure the power supply of the socket after it is normal Replacing the socket position, the fault persists,
2. When testing the fixing component with a normal and good machine of the same model, the fault still reappeared
3. The fault was not resolved after swapping with the normal machine power board
4. Problem solved after replacing BICU board